Class I recall: 3B Medical Luna G3 bilevel positive airway pressure devices record set pressure instead of delivered pressure when Targeted VT is on

The Food and Drug Administration (FDA) recall database posted a Class 1 recall of the 3B Medical Luna G3 BPAP (Bi-level Positive Airway Pressure) System on November 12, 2026. When the Targeted VT feature is enabled, the device records and transmits the configured inspiratory pressure value instead of the pressure actually delivered. Customers are to stop using Targeted VT, remove affected devices from clinical service and segregate them, and use unaffected devices for patient support until a firmware correction is installed. The record lists 12,672 units distributed nationwide.

Background

The Luna G3 is a bilevel positive airway pressure device for treating adult Obstructive Sleep Apnea (OSA). The firm began notifying customers on October 9, 2026.

Who this affects

Sleep laboratories, respiratory therapy and durable medical equipment programs that use or dispense Luna G3 BPAP System model LG3800-30VT

Why it matters

When Targeted VT is on, affected devices record and transmit the configured pressure, not the pressure delivered, and the devices are to be taken out of service until corrected, which may require substitute devices for patients. Inference: clinicians reviewing downloaded therapy data from these devices may have been seeing set rather than delivered pressures.

Required actions

  • None stated as a legal requirement. The recall record's Action field gives the manufacturer's recall instructions: customers should immediately discontinue use of Targeted VT if enabled on affected devices, affected devices should be removed from clinical service and segregated to prevent further use, customers should use unaffected positive airway pressure devices for patient support, and the firmware correction will be sent upon receipt of the completed customer response form.

Scope and timing

Authority status
Other (FDA Class 1 recall record)
Responsible owner
Respiratory therapy, Sleep medicine, Clinical engineering, Durable medical equipment
Publication date
2026-11-12
Other dates
2026-10-08 recall initiated by firm; 2026-10-09 customer notification began

Not established

No FDA recall notice page or early alert was found for this recall; the facts come from the recall database record.

Sources

Complete edition and supplemental material

This article in the complete report →

Executive triage, source-coverage limitations, context and forward horizons remain in the complete report.

Read the entire Compliance Sweep report →

← Return to stories

Privacy preferences

Where enabled for this site, Cloudflare Web Analytics can send page-use and performance measurements to Cloudflare. This panel provides information and does not turn analytics on or off.

This panel stores no tracking choice and does not change your tool saves, screening, resets, reports or exports. Other sites have separate settings.

Read Privacy & Cookies